Flow rate:1.5~30L/min, Pressure in :4.5-10 bar
Model: GC1500 series
Order NO.:
405
Specific parameters will be adjusted according to product upgrades, without notice.
Chemtron GC Zero Air Generators are designed to supply ultrapure hydrocarbon-free zero air for GC-FID applications. Catalytic oxidation technology eliminates hydrocarbons to <0.1ppm, resulting in a flat, stable baseline, increased sensitivity, and repeatable analysis.
GC Zero Air features:
* Flow rate from 1.5 to 30 l/min
* Pressure up to 6.5 bar (95 psig)
* Zero air purity <0.1ppm HC
* Minimal maintenance
* 2 year standard warranty
Operating principle
Zero Air Generators use three steps to transform ambient air into analytical grade air.
STEP 1: PRE-FILTRATION
An external oil-free compressor delivers air through a high efficiency filter that traps any particles that may damage the system. The filter, which has an automatic purge system, traps oil, water, and any particles larger than 5 microns in size.
STEP 2: HC AND CO TRAPPING
The air leaving the filter enters a high temperature platinum catalyzer, which eliminates hydrocarbon molecules down to < 0.1 ppm through oxidation.
STEP 3: FINAL FILTRATION
A final pass through a high-efficiency filter prevents any kind of particles from entering the instrument.
Benefits
Better detector performance
The reduction of carbon monoxide and hydrocarbons (including methane) to < 0.1 ppm decreases the background noise level and gives the baseline much better stability, increasing detector sensitivity considerably and ensuring precise analytical results.
Easy installation
Just connect a suitable oil-free compressor (optional). To save bench space, the unit can be easily wall-mounted.
Lab safety
No more bottles in the lab and expensive pipelines for air distribution.
Save money
All that the unit requires is a connection to a suitable oil-free compressor and the mains. Typically the investment is recovered in less than a year.
Improved efficiency
Round the clock continuous operation maximizes lab productivity, eliminating dead time for gas bottle changeover and avoiding the need for tedious instrument recalibrations.
Models |
GC1500 |
GC3000 |
GC6000 |
GC15000 |
GC30000 |
Flow ( L/min ) |
1.5 |
3 |
6 |
15 |
30 |
Weight ( kg ) |
5 |
9 |
12 |
22 |
22 |
HC & CO out |
< 0.1 |
||||
Max CO in |
50 ppm |
||||
Max HC in |
100 ppm |
||||
Max temp. in |
40 ℃ |
||||
Pressure in |
4.5-10 barb Additional air compression sources are required |
||||
Pressure drop |
<1 bar |
||||
Stability ( min ) |
45 |
||||
In connections |
1/4 ( NPT ) |
||||
out connections | 1/8 ( NPT ) | ||||
Power |
110-120V 60Hz / 220-240V 50Hz |
||||
Dimensions ( mm ) |
410H×550W×230D |
||||
Working temp |
Amb.+15℃ |
||||
Certification |
CE, CSA, FCC |